Abstract

Sophisticated sample preparation including the determination of stable nuclides are an essential prerequisite for high-accuracy accelerator mass spectrometry (AMS) data. Improvements in the low-level regime already paid back [1], however, some pitfalls still exist or are (re-) appearing due to recent developments:

1.) As most samples prepared for 10Be-AMS need the addition of 9Be in the form of a liquid solution of known 9Be-concentration and commercial solutions contain too much 10Be, solutions from minerals originating from deep mines have been established. Special attention has recently been paid to the preparation of such a 9Be-carrier by the determination of the 9Be-value by an interlaboratory comparison. It could be shown that deviations between different labs exist, thus, it is strongly advised to have such solutions analysed at more than a single lab to prevent incorrect 10Be-results.
2.) In our approach to analyse as many radionuclides as possible in a single meteorite sample, small changes in the established chemical separation [2] have been tested. Though, the secondary formation of partially insoluble compounds of Mg and Al by the pressure digestion is strongly influenced, thus, yielding to too low 27Al-results in the taken aliquot and overall incorrect 26Al-results.
Ackn.: DREAMS-team and colleagues performing 9Be- and 27Al-measurements (BAM, GFZ, HZDR, CEREGE, TUBAF, U Cologne, VKTA). Ref.: [1] NIMB 266 (2008) 4921. [2] RCA 84 (1999) 215.

Keywords: AMS; stable nuclides; radionuclides

Beteiligte Forschungsanlagen

Verknüpfte Publikationen

  • Vortrag (Konferenzbeitrag)
    DPG-Frühjahrstagung der Sektion AMOP (SAMOP), 18.-22.03.2013, Hannover, Deutschland

Downloads

Permalink: https://www.hzdr.de/publications/Publ-18131