Comments on “Comparative Study With New Accuracy Metrics for Target Volume Contouring in PET Image Guided Radiation Therapy”


Comments on “Comparative Study With New Accuracy Metrics for Target Volume Contouring in PET Image Guided Radiation Therapy”

van den Hoff, J.; Hofheinz, F.

Abstract

Phantom measurements with glass inserts in a hot background are still used frequently for calibration and performance assessment of volume delineation algorithms in PET. Taking as an example the recent paper by Shepherd et al.: Comparative Study with New Accuracy Metrics for Target Volume Contouring in PET Image Guided Radiation Therapy (IEEE Trans Med Imaging. 2012 Jun 4. [Epub ahead of print]), we demonstrate that this is not a valid approach due to the discontinuity introduced in the background by the cold walls of the glass inserts. We, moreover, emphasize that in order to define a sensible ground truth for performance assessment of contouring algorithms in patient data it is necessary to average over a sizable number of experienced observers and lesions in order to compensate for the substantial inter-observer variability.