Computer assisted beam modeling for particle therapy


Computer assisted beam modeling for particle therapy

Fuchs, H.; Elia, A.; Resch, A. F.; Kuess, P.; Lühr, A.; Vidal, M.; Grevillot, L.; Georg, D.

Abstract

Purpose:

To develop a computer-driven and thus less user dependent method, allowing for a simple and straight forward generation of a Monte Carlo (MC) beam model of a scanned proton and carbon ion beam delivery system.

Methods:

The method was applied on five different clinical as well as one research beam lines for proton and carbon ions of three different particle therapy centers using synchrotron or cyclotron accelerator systems: i) MedAustron ion therapy center, ii) University Proton Therapy Dresden, and iii) Center Antoine Lacassagne Nice. In a first step, experimental measurements were performed for proton and carbon ion energies in the available energy ranges. Data included depth dose profiles measured in water and spot sizes in air at various iso-center distances.
Using an automated regularization-based optimization process, GATE/Geant4 beam models of the respective beam lines were generated and compared to independent measurements. Sequentially, using least square weighting functions with and without regularization, the beam parameters energy, energy spread, beam sigma, divergence, and emittance were iteratively tuned until a user defined agreement was reached. Based on the parameter tuning for a set of energies a beam model was generated. The resulting beam models were validated for all centers comparing laterally integrated depth dose curves and spot sizes in air. For a representative center, 3D dose cubes were measured and compared to simulations.

Results: Beam ranges in the MC beam models agreed on average within 0.2 mm compared to measurements for all energies and beam lines. Spot sizes (full-width at half maximum) at all positions in air, differed by less than 0.4% from the measurements.
Dose calculation with the beam model for the MedAustron clinical beam line agreed better than 1.7% in absolute dose for a representative clinical case treated with protons.
For protons, beam model generation, including geometry creation, data conversion, and validation, was possible within three working days.The number of iterations required for the optimization process to converge, was found to be similar for all beam line geometries and particle types.

Conclusion:

The presented method was demonstrated to work independently of the beam optics behavior of the different beam lines, particle types and geometries. Furthermore, it is suitable for non-expert users and requires only limited user interaction. Beam model validation for different beam lines based on different beam delivery systems, showed good agreement.

Keywords: Monte Carlo; beam modeling; optimization; proton therapy; carbon ions; particle therapy